R Series Service Manual
Disassembling the System Brick Assembly
Note:
See the safety precautions on page 3-2 before removing the system brick assembly.
The system brick assembly consists of three primary boards attached together, Units equipment with SpO
boards.
Top: Digital system board (SpO
2
Middle: Analog system board
Bottom: Pace defibrillator engine
Use caution when pulling apart the three boards to prevent the EMI suppression plates from scattering.
Removing Isolated Power Supply
1. Remove 5 screws holding the isolated power supply.
Note:
During reassembly torque screws to 4 in-lbs.
3-28
module and isolated power supply sits on digital board)
and/or EtCO
2
2. Peel back glue from connector, and disconnect cable
by pulling straight up. Remove tape holding cable to
isolated power supply, then disconnect cable.
Note:
During reassembly re-apply glue to each side of connector.
& NIBP will have two additional
2
9650-0903-01 Rev. L