Press the Manual Mode softkey on the front panel of the unit to enter the Manual mode of
operation.
The message CONFIRM MANUAL MODE and the Confirm softkey will be displayed. Press
the Confirm softkey to enter the Manual mode of operation. If the Confirm softkey is not
pressed within 10 seconds, the unit will revert back to semiautomatic operation.
Note:
The control buttons for Manual mode light up on the front panel.

2 Turn Mode Selector to PACER

RECORDER
ON
OFF
PACER
The pacer door drops open and the knobs are displayed.
Set the Pacer Output to 0 mA
If the unit has just been turned on, the PACER OUTPUT is automatically set to 0 mA.

3 Set Pacer Rate

Set the PACER RATE to a value 10-20 ppm higher than the patient's intrinsic heart rate. If no
intrinsic rate exists, use 100 ppm.
The pacer rate increments or decrements by a value of 2 ppm on the display when the knob is
turned.
OFF
PACER
OUTPUT
OUTPUT
9–4
T
P
EMPORARY
ACING
If the unit has been configured not to confirm the transition from AED mode to
Manual mode, the unit will immediately enter Manual mode.
1
4:1
RATE
RATE
mA
mA
ppm
ppm