R Series Service Manual
8.0 Calibrating Pulses on Strip Chart Test
Tools Needed: None.
Test Setup: None
.
Do this...
8.1
Press the RECORDER button.
8.2
Press and hold SIZE button to activate the
calibration signal.
Record your results on the Maintenance Test Checklist.
1-18
Observe this...
The strip chart displays a signal of 300 ppm with an amplitude of 10 mm
+/- 1 mm. The signal also appears on the video display. (You can verify that the
rate is 300ppm by measuring 5mm from the left edge of one pulse to the left
edge of the following pulse.)
Pass/Fail/NA
o o o
9650-0903-01 Rev. L