11.0 Defibrillator Self Test

SHOCK HAZARD!
TAKE THE NECESSARY PRECAUTIONS TO GUARD AGAINST SHOCK OR INJURY BEFORE YOU START
CONDUCTING THE DEFIBRILLATOR TESTS.
Keep hands and all other objects clear of the multi-function cable connections and defibrillator analyzer when discharging
the defibrillator.
Before you discharge the defibrillator, warn everyone near the equipment to STAND CLEAR.
Caution
Do NOT internally discharge the unit more than 3 times in 1 minute. Note that multiple rapidly repeating internal discharges at more than
30 Joules may damage the unit.
Tools Needed:
ZOLL Medical Electrode Connector (Fluke Biomedical Part Number 3010-0378 or equivalent).
Impulse 4000 or equivalent defibrillator analyzer.
ECG Cable.
Stop watch.
Test Setup:
1. Ensure the unit is turned off.
2.
Ensure the ECG cable is connected to the R Series unit and the defibrillator analyzer.
Note:
The OneStep cable should not be connected to any equipment at the beginning of this test.
Do this...
11.1
Turn the selector switch to DEFIB mode (For
AED units, turn to ON, and select MANUAL
MODE.)
Set leads to PADS.
11.2
Connect the OneStep cable to the test port on
the right side of the R Series.
1-22
Observe this...
CHECK PADS/POOR PAD CONTACT message displays.
DEFIB PAD SHORT message displays.
Pass/Fail
o o
o o
9650-0903-01 Rev. L