15.0 Read Error Log - Final

Equipment Needed
Power Supply 12V 10A,
Battery Current Test Fixture (P/N 9100-0690-TF) or AED PRO Disposable Battery (P/N
1008-002003-01),
AED PLUS Quick-Test Test Fixture (P/N 9100-0567-TF),
Setup:
Start the ZAS on the computer; make sure the connection is established before starting the
AED Pro unit.
Note: For best transmission results, the ZOLL USB IrDA port (P/N 9355-0807) must be
facing the IrDa port of the unit with a clear 10 to 18-inch line-of-sight.
Connect AED PLUS Quick-Test Test Fixture (P/N 9100-0567-TF) to the unit.
Test
15.1
15.2
15.3
15.4
1–22
T
ESTS
Action
If the unit is on, press the On/Of
button for 1 second to turn off
the unit. .
Press and hold the On/Off
button for at least 7 seconds,
then release it.
Double click on the latest ZAS
AED Pro version software icon
on Desktop.
Click on History and select
"upload and display history data
from device".
Click NEXT and notice "*.CEH
" in the filename box and click
SAVE.
Click on the ERROR LOG box.
Clear Log and Close the AED
Pro Error Log box.
www.zoll.com
Expected Result(s)
Unit powers off.
The AED Pro unit establishes
contact with the Computer
within 5 seconds and the unit
issues the voice and text prompt
message UNIT OK, followed
by NONRESCUE MODE and
COMMUNICATION
ESTABLISHED
Verify ERROR ID codes and
ERROR DESCRIPTIONS.
Pass
Fail
9650-0309-01 Rev. E