Manual Mode Activated

AED versions of the E Series will record a "MANUAL
MODE STARTED" summary report event when the
device is switched from Semiautomatic mode to Manual
mode.

Printing a Report

To print the stored information, press the SUMMARY
button below the screen display. Then press the
corresponding softkey to print a "call" report, summary
report (print chart), or incident log.
Print
Print
Print
Call
Chart
Log
The recorder prints all summary report events currently
in memory in chronological order. If the unit is equipped
with the 12-lead option, the recorder will print all 12-lead
patient records currently in memory at the end of the
summary report printout. If the stripchart recorder is on
or the defibrillator is charged, summary report printing is
disabled. In addition:
To stop printing a report, press the RECORDER
button or turn the unit off. You can print an unlimited
number of copies of the report by pressing the
SUMMARY button and corresponding print softkey
again.
If you press the RECORDER button while printing a
summary report, the unit stops printing the report.
Press the RECORDER button again to begin printing
an ECG trace. The stripchart recorder runs
continuously until the button is pressed again.
If you press the SUMMARY button and a
corresponding print softkey while a report is already
printing, the current report stops printing and a new
report begins to print.
Printing is interrupted if a vital sign alarm occurs (i.e.
HR, SpO2, etc.), the ANALYZE button is pressed, or
the defibrillator is charged.
If the recorder is out of paper when the SUMMARY
button and a corresponding print softkey are pressed,
a CHECK RECORDER message appears on the
display. Load paper and press the SUMMARY button
again to select the report to print.

Printing a Call Report

A Call Report is an abbreviated summary report
containing only those events associated with a specific
call or run. To print a call report:
1. Press the SUMMARY button.
2. Press the Print Call softkey.
9650-1210-01 Rev. R
Realtime
Trend
Start
Operating Controls and Indicators
3. Use the scroll keys on top of the unit to scroll through
the list of start times associated with different calls.
---
NIBP
SpO2%
kPa
24
---
---
-
---
CO2
m
---
Hg
m
02NOV 11:43:11 Call Started
02NOV 11:40:05 Call Started
02NOV
39 52 Call Started
02NOV 11:38:11 Call Started
02NOV 11:35:05 Call Started
Print
Records
4. Press the Commit ( ) key on top of the unit, or the
Print Records softkey, to print the events associated
with the selected call.

Printing a Partial Summary Report

To print only a portion of the Summary Report:
1. Press the SUMMARY button.
2. Press the Print Chart softkey.
3. Press the Print Range softkey.
4. Use the scroll keys on top of the unit to scroll through
the list of events.
---
NIBP
SpO2%
kPa
24
---
---
-
---
m
CO2
---
Hg
m
02NOV 14:18:02 Intubate
02NOV 14:12:29 CPR
02NOV 11:54:03 DEFIB ADVISORY
02NOV 11:52:48 RECORDED ECG
02NOV 11:06:14 POWER ON
Print
Records
5. Press the Commit ( ) key on top of the unit, or the
Print Records softkey, to print the specified event
and all subsequent events.

Printing an Incident Log

An incident log is an abbreviated list of all major events
recorded in the summary report. You can print out an
incident log that includes the time of occurrence of the
following events:
E Series unit powered on
Defibrillation advisory messages (for example,
CHECK PATIENT and SHOCK ADVISED)
Defibrillation shocks (including energy level)
Pacer mode activated
PADS
x1
ECG
70
RR
Abort
ECG
PADS
x1
70
RR
Abort
2-9

Hide quick links: