10.8 Test Mode - Rice Lake IQ plus 710 Installation Manual

Digital weight indicator
Hide thumbs Also See for IQ plus 710:
Table of Contents

Advertisement

10.8 Test Mode

In addition to normal and setup modes, test mode
provides a number of diagnostic functions for the IQ
plus 710, including:
Display raw A/D count
Set digital outputs on and display digital input
states
Reset configuration parameters to default
values
Clear non-volatile (battery backed) storage
Clear EEPROM
Discontinued Product
Set analog output state to zero or full scale
Set A/D offset and gain calibration
DI/O1=
DI/O2=
DI/O5=
DI/O6=
AOUT
=0
TEST Menu
Function
DI/O1 —DI/O8
Set digital output x high and display status of digital input x
Press and hold DI/Ox= key (see Figure 10-6) to show status of DIGINx (DIx=HI or DIx=LO).
AOUT=0
Set analog output to zero
Press and hold the F1 key to set analog output to its zero value.
AOUT=FS
Set analog output to full scale
Press and hold the F2 key to set analog output to its full scale value.
A/D TEST
Display A/D test
Press and hold the 7 key to display raw count from A/D converter.
CLEAR NVRAM Clear non-volatile storage
Press and hold the setup switch, then press the ENTER key to clear values stored in battery-backed SRAM,
including truck mode data, time, and date.
DEFAULT
Default parameters
Press and hold the setup switch, then press the 0 key to reset configuration and calibration parameters to
factory default values. Load cells must be recalibrated before using the indicator (see Section 4.0 on
page 37).
72
IQ plus 710 Installation Manual
DI/O3=
DI/O4=
DI/O7=
DI/O8=
AOUT
=FS
Figure 10-6. Front Panel Key Functions in Test Mode
Table 10-5. Test Menu Functions
To enter test mode, press and hold the setup switch
until the front panel display shows the word
test mode, the front panel keys are mapped to the test
mode functions as shown in Figure 10-6 and
Table 10-5.
A/D calibration functions, ADOFFS and
Caution
ADGAIN, must be used only by qualified
service
replacing
A/D
converter
calibration may render the indicator unusable.
The CLEAR EEPROM function erases both A/D and load cell
calibration data. The A/D converter and the scale must be
recalibrated after using this function.
A/D
A/D
TEST
OFFSET
CLEAR
DEFAULT
EEPROM
Description
. In
TEST
personnel,
and
only
after
components.
Improper
A/D
EXIT
A/D
GAIN
CLEAR
NVRAM

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents