3.0

Configuration

To configure the IQ plus 710 indicator, the indicator
must be placed in setup mode. The setup switch is
accessed by removing the large fillister head screw on
the enclosure backplate. Switch position is changed
by inserting a screwdriver into the access hole and
pressing the switch.
When the indicator is placed in setup mode, the word
is shown on the display. The CONFIG menu
CONFIG
is the first of ten main menus used to configure the
indicator. Detailed descriptions of these menus are
Discontinued Product
given in Section 3.2. When configuration is complete,
return to the CONFIG menu and press the
key to exit setup mode, then replace the setup switch
access screw.
3.1

Configuration Methods

The IQ plus 710 indicator can be configured by using
the front panel keys to navigate through a series of
configuration menus or by sending commands or
configuration data to the EDP port. Configuration
using the menus is described in Section 3.1.3.
C o n f i g u r a t i o n u s i n g t h e E D P p o r t c a n b e
accomplished using either the EDP command set
described in Section 5.0 or Version 2.5 or later of the
Revolution
configuration utility.
3.1.1

Revolution Configuration

The Revolution configuration utility provides the
preferred method for configuring the IQ plus 710
indicator. Revolution runs on a personal computer to
set configuration parameters for the indicator. When
Revolution configuration is complete, configuration
data is downloaded to the indicator.
Figure 3-1. Sample Revolution Configuration Display
14
IQ plus 710 Installation Manual
Revolution supports both uploading and downloading
of indicator configuration data. This capability allows
configuration data to be retrieved from one indicator,
edited, then downloaded to another.
To use Revolution, do the following:
1. Install Revolution (Version 3.0 later) on an
IBM-compatible personal computer running
Windows
Minimum system requirements include a
processor speed of at least 133MHz, 32MB of
memory (64MB recommended), and at least
(
)
ZERO
20MB of available hard disk space for
installation.
2. With both indicator and PC powered off,
connect the PC serial port to the RS-232 pins on
the indicator EDP port.
3. Power up the PC and the indicator. Use the setup
switch to place the indicator in setup mode.
4. Start the Revolution program.
Figure 3-1 shows an example of one of the Revolution
configuration displays.
Revolution provides online help for each of its
configuration displays. Parameter descriptions
provided in this manual for front panel configuration
can also be used when configuring the indicator using
Revolution: the interface is different, but the parameters
set are the same.
3.1.2

EDP Command Configuration

The EDP command set can be used to configure the
IQ plus 710 indicator using either a personal
computer, terminal, or remote keyboard. Like
Revolution, EDP command configuration sends
commands to the indicator EDP port; unlike
Revolution, EDP commands can be sent using any
external device capable of sending ASCII characters
over a serial connection.
EDP commands duplicate the functions available
using the indicator front panel and provide some
functions not otherwise available. EDP commands can
be used to simulate pressing front panel keys, to
configure the indicator, or to dump lists of parameter
settings. See Section 5.0 on page 41 for more
information about using the EDP command set.
®