Calibration Menu - Rice Lake IQ plus 710 Installation Manual

Digital weight indicator
Hide thumbs Also See for IQ plus 710:
Table of Contents

Advertisement

3.2.3

Calibration Menu

See Section 4.0 on page 37 for calibration procedures.
CONFIG
XXXXXXX
FORMAT
WZERO
*CALIB*
Display and edit
Display and edit
test weight value
zero calibration
A/D count value
Discontinued Product
PT–> 1
Display and edit
test weight value
*CALIB*
Display and edit
linearization point
A/D count value
CALIBR Menu
Parameter
Choices
Level 2 submenus
WZERO
Press ENTER to display and edit the zero calibration A/D count value.
WVAL
Press ENTER to display and edit the test weight value.
WSPAN
Press ENTER to display and edit the span calibration A/D count value.
WLIN
PT->1 —
Press ENTER to display and edit test weight and calibration values for up to five linearization points.
PT->5
Perform linear calibration only after WZERO and WSPAN have been set.
REZERO
Press Enter to remove an offset value from the zero and span calibrations.
Use this parameter only after WZERO and WSPAN have been set. See Section 4.1 on page 37 for
more information about using this parameter.
22
IQ plus 710 Installation Manual
CALIBR
SERIAL
PROGRM
WVAL
WSPAN
*CALIB*
Display and edit
span calibration
A/D count value
PT–> 2
PT–> 3
Same as PT-> 1
Figure 3-6. Calibration Menu
Table 3-4. Calibration Menu Parameters
XXXXXXX
PFORMT
SETPTS
XXXXXXX
DIGIN
WLIN
REZERO
CALIB*
*
Press Enter to
remove offset from
zero and span
calibrations
PT–> 4
PT–> 5
Description
XXXXXXX
ALGOUT
VERS

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents