TEST Menu
Function
CLEAR EEPROM Clear EEPROM
Read Caution! statement on page 72 before using this function.
Press and hold the setup switch, then press the
A/D OFFSET
A/D offset calibration (–0.5 mv/V)
Read Caution! statement on page 72 before using this function. Press and hold the setup switch, then
press the 8 key to perform offset calibration.
A/D GAIN
A/D gain calibration (+4.5 mv/V)
Read Caution! statement on page 72 before using this function. Press and hold the setup switch, then
press the 9 key to perform gain calibration.
Discontinued Product
EXIT
Press the 3 key to exit test mode.

10.9 Software Revision Information

Version 2.1 of the IQ plus 710 software provides the
following additional support:
New SMPRAT parameter on the CONFIG
menu provides selectable A/D conversion rate
ranging from 3.75–30 Hz.
New YYMMDD value on the DATEFMT
parameter (PROGRM menu).
Description
.
Table 10-5. Test Menu Functions (Continued)
(decimal_point) key to clear EEPROM.
Appendix
73