Revolution Calibration - Rice Lake IQ plus 710 Installation Manual

Digital weight indicator
Hide thumbs Also See for IQ plus 710:
Table of Contents

Advertisement

4.3
Revolution
Calibration
To calibrate the indicator using Revolution, the
indicator EDP port must be connected to a PC running
the Revolution configuration utility.
Use the following procedure to calibrate the indicator:
1. Place the indicator in setup mode (display
reads
) and remove all weight from the
CONFIG
scale platform.
2. Select
Calibrate Indicator
main menu.
3. On the Indicator Calibration display, select
t h e i n d i c a t o r m o d e l (
Discontinued Product
communications port then click
4. Revolution uploads calibration data from the
indicator then presents the information in a
display like that shown in Figure 4-2.
Figure 4-2. Revolution Calibration Display
from the Revolution
) a n d
I Q + 7 1 0
.
OK
5. Enter the
Value of Test Weight
span calibration then click
6. A dialog box asks whether hooks or chains
are used to hang calibration test weights.
Click
or
to continue.
Yes
No
7. The Zero Calibration dialog box prompts you
to remove all weight from the scale. Clear the
scale and click
to begin zero calibration. If
OK
your test weights require hooks or chains,
place the hooks or chains on the scale for zero
calibration.
8. When zero calibration is complete, the Span
Calibration dialog box prompts you to place
test weights on the scale for span calibration.
Place test weights on the scale then click
9. When span calibration is complete, a dialog
box asks whether you wish to perform linear
calibration. Click
calibration for up to five linearization points, or
click
to continue calibration with step 11.
No
10. On the Linear Calibration display, select the
point (1–5) to calibrate, then click
Place test weights on scale then click
When prompted, enter the test weight value
then press Enter. Repeat for additional
linearization points, then click
the Indicator Calibration display.
11. If hooks or chains were used to hang test
weights, the Rezero dialog box prompts you
to remove weights from the scale. Remove
the weights then click
If hooks or chains were not used in the
calibration procedure, Revolution goes directly
to step 12. Calibration is complete.
12. When calibration is complete, the
fields of the Indicator Calibration display are
filled in. Click
to save the new values and
Exit
return to the Revolution main menu; to restore
the previous calibration values (including
linear calibration values), click
.
Settings
to be used for
.
START
.
OK
to perform linear
Ye s
.
Calibrate
.
OK
to return to
Exit
.
OK
New Settings
Restore
Calibration
39