Calibration; Front Panel Calibration - Rice Lake IQ plus 710 Installation Manual

Digital weight indicator
Hide thumbs Also See for IQ plus 710:
Table of Contents

Advertisement

4.0

Calibration

The IQ plus 710 can be calibrated using the front panel, EDP commands, or the Revolution
utility. Each method consists of the following steps:
Zero calibration
Entering the test weight value
Span calibration
Optional five-point linearization
Optional rezero calibration for test weights using hooks or chains.
The following sections describe the calibration procedure for each of the calibration methods.
Discontinued Product
CONFIG
XXXXXXX
FORMAT
WZERO
WVAL
*CALIB*
Display and edit
Display and edit
test weight value
zero calibration
A/D count value
PT–> 1
Display and edit
test weight value
*CALIB*
Display and edit
linearization point
A/D count value
4.1

Front Panel Calibration

To calibrate the indicator using the front panel, do the
following:
1. Place the indicator in setup mode (display
reads
) and remove all weight from the
CONFIG
scale platform. If your test weights require
hooks or chains, place the hooks or chains on
the scale for zero calibration.
2. Press
until the display reads
Figure 4-1). Press
(
).
WZERO
3. With
displayed, press
WZERO
zero. The indicator displays
calibration is in progress. When complete, the
A / D c o u n t f o r t h e z er o c a l i b r a t i o n i s
displayed. Press
calibration value and go to the next prompt
(
).
WVAL
CALIBR
SERIAL
PROGRM
WSPAN
*CALIB*
Display and edit
span calibration
A/D count value
PT–> 2
PT–> 3
Same as PT-> 1
Figure 4-1. Calibration (CALIBR) Menu
(see
CALIBR
to go to zero calibration
to calibrate
while
*CAL*
again to save the zero
XXXXXXX
PFORMT
SETPTS
XXXXXXX
DIGIN
WLIN
REZERO
CALIB*
*
Press Enter to
remove offset from
zero and span
calibrations
PT–> 4
PT–> 5
4. With
displayed, place test weights on
WVAL
the scale and press
value. Use the numeric keypad to enter the
actual test weight, then press
the value and go to span calibration (
5. With
displayed, press
WSPAN
span. The indicator displays
calibration is in progress. When complete, the
A / D c o u n t f o r t h e s p a n c a l i b r a t i o n i s
displayed. Press
calibration value and go to the next prompt
(
).
WLIN
6. Five-point linearization (using the WLIN
parameter) provides increased scale accuracy
by calibrating the indicator at up to five
additional points between the zero and span
calibrations.
configuration
XXXXXXX
ALGOUT
VERS
to show the test weight
to save
ENTER
).
WSPAN
to calibrate
while
*CAL*
again to save the span
Calibration
37

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents