10.7 Analog Output Calibration

The following calibration procedure requires a
multimeter to measure voltage or current output from
the analog output module. If the option is not already
installed, see Section 2.4 on page 7.
CONFIG
XXXXXXX
FORMAT
CALIBR
SOURCE
OFFSET
0%
GROSS
20%
NET
1. Enter setup mode and go to the ALGOUT
menu (see Figure 10-5):
Set OFFSET to 0% for 0–10 V output,
20% for 4–20 mA output
Set MIN to lowest weight value to be
tracked by the analog output
Set MAX to highest weight value to be
tracked by the analog output
2. Connect multimeter to connector J1 on the
analog output board:
For voltage output, connect voltmeter
leads to pins 3 and 4
For current output, connect ammeter
leads to pins 1 and 2
Secondary Unit
grains
grams
kilograms
ounces
pounds
troy pounds
NOTE: The analog output must be calibrated after the
indicator itself has been configured (Section 3.0) and
calibrated (Section 4.0).
SERIAL
PROGRM
PFORMT
XXXXXXX
ERRACT
MIN
000000
FULLSC
HOLD
number
ZEROSC
Figure 10-5. Analog Output Menu
Primary Unit
x Multiplier
troy pounds
5760
373.2417
0.373242
13.16571
0.822857
12
Table 10-4. Conversion Factors (Continued)
SETPTS
XXXXXXX
DIG IN
ALGOUT
XXXXXXX
MAX
10000
number
3. Adjust zero calibration: Scroll to the
TWZERO parameter. Press
value, then check voltage or current reading
on multimeter. Press and hold
adjust the zero value up or down.
4. Adjust span calibration: Scroll to the
TWSPAN parameter. Press
value, then check voltage or current reading
on multimeter. Press and hold
adjust the span value up or down.
5. Final
zero
calibration:
TWZERO parameter and verify that the zero
calibration has not drifted. Press and hold
or
to re-adjust the zero value as required.
6. Return to normal mode. Analog output
function can be verified using test weights.
Secondary Unit
grains
grams
kilograms
ounces
pounds
troy ounces
XXXXXXX
VERS
TWZERO
TWSPAN
71
15171
number
number
to view zero
or
to
to view span
or
to
Return
to
the
Appendix
71