Master Power Up Screen - KEPCO BOP 1000W Operator's Manual

High power bipolar power supply
Hide thumbs Also See for BOP 1000W:
Table of Contents

Advertisement

a. From the power-up screen press
Parallel and press
prompted, enter ADMIN2 password (see PAR. 3.2.4.4) before continuing.
b. From the Multiple Units menu, highlight Configuration (use ADJUST control or the
U
or
keys) and press
3X2, press
c. If Series or Parallel was selected in step b above, Connection
Series or Parallel as selected above. Highlight Unit Type and press
Highlight Master +1
ter +4 not permitted for series connections) and press
d. Exit the Multiple Units menu by pressing
cycles. The display shows the master power-up screen (see Figure 2-9).
e. Turn off power to the master, then turn on power to the master.
f. The power-up master module screen (Figure 2-9) shows the status of the slave(s). Refer
to Table 2-13 for an explanation of status indications for the slave(s).
If all the slaves are recognized as ready, after about 10 seconds the master LCD reverts to the
Power On Default screen (see Figure 2-10). The slaves display the Slave Power On Default
screen (see Figure 2-11), except the unit will be in Current Mode for parallel configurations and
Voltage mode for series configurations and the message will indicate the unit is operating as a
slave. The slave number appears at the lower left of the display. For instructions to return the
slave to standalone operation, refer to PAR. 2.8.5. The fault indicators on all units are off.
BOP HIPWR 080709
%
!
to view the Multiple Units menu (see Table 2-12). When
!
to modify. Select Series, Parallel, Master 2X2 or Master
$
to apply the change.
,
Master +2
FIGURE 2-9.

MASTER POWER UP SCREEN

to enter the General Setup Menu. Highlight Series/
,
, or
Master +4 (Master +3
Master +3
$
$
to save the settings for subsequent power-up
Type shows
!
to modify.
, or
to apply the change.
Y
Mas-
2-21