List:coun T :Skip Command - KEPCO BOP 1000W Operator's Manual

High power bipolar power supply
Hide thumbs Also See for BOP 1000W:
Table of Contents

Advertisement

.VOLT:MODE LIST
VOLT?
LIST:COUNT?
LIST:COUNT 10
LIST:COUNT?
VOLT:MODE LIST
VOLT?
LIST:DWELL .01,.01,.01,.01,.01,.01,.01,.01,.01
LIST:DWELL .03
VOLT:MODE LIST
LIST:DWELL:POINTS?
LIST:VOLT:POINTS?
LIST:DWELL .01,.01,.01,.01,.01,.01,.01,.01,.01,.01
DWELL:POINTS?
LIST:QUER?
LIST:QUER 9
LIST:DWEL?
LIST COUNT 0
VOLT:MODE LIST
LIST:DWEL?
VOLT:MODE FIX
LIST:VOLT:APPLY SINE,60,20 Adds a 60Hz, 20V p-p sine wave to the current waveform
VOLT:MODE:LIST
*RST
FIGURE B-5.
B.50 [SOURce:]LIST:COUNt:SKIP COMMAND
Syntax:
Short Form: LIST:COUN:SKIP nn
nn = <int_value 0 to 255>
Description: Allows beginning steps of list-generated waveform to be run once, then ignored.
When a list is to be repeated using LIST:COUNT, this command allows the user to skip the first nn
steps once the full set has been executed. After the first iteration (which executes all steps), the first
nn steps are skipped. The LIST:COUN:SKIP command allows the user to precondition a list-generated
waveform by setting unique conditions at the beginning that are not repeated for the rest of the repeti-
tions. LIST:CLEar sets nn to 0. Related Commands: LIST:COUN, LIST:COUN:SKIP?, LIST:CLE. (See
example, Figure B-5.)
B.51 [SOURce:]LIST:COUNt:SKIP? QUERY
Syntax:
Short Form: LIST:COUN:SKIP?
Return Value: <int_value>
Description: Identifies how many steps will skipped the first time the list is executed. Returns value set by
LIST:COUN:SKIP command. (See example, Figure B-5.)
B-14
Executes the list. For 240 milliseconds the BOP outputs a staircase triangle
wave from -20V to +20V and back down to -20V. This staircase will have a
uniform spacing between voltage changes of 10 milliseconds and will
repeat 100 times.
Returns +20 (the last step in the list set the unit to +20V.
Returns 100.
Determines that the list will be repeated 10 times when executed.
Returns 10.
Initiates execution of the list. The unit outputs a sawtooth waveform that
increases from -20V to +20V and repeats 10 times. There is an abrupt
change from +20V to -20V each time the list transitions from location 20 to
location 0.
Returns 20, (the last step in the list set the unit to 20V.
Enters dwell time of 0.01 seconds in locations 1 through 9.
Enters dwell time of 0.03 second in location 10.
the list is not executed, the unit returns error -221,Settings Error indicating
the lists are not balanced.
Returns 11 to indicate the dwell list has 11 entries
Returns 21 to indicate the voltage list has 21 entries.
Dwell times are entered in locations 11 through 20.
Returns 21 indicating there are now 21 step dwell times on the list.
Returns 0 indicating list queries will start from location 0
Causes list queries to start at location 9
Returns 01,.03,.01,.01,.01,.01,.01,.01,.01,.01,.01,.01(the dwell times for
locations 9 through 20.
List will repeat indefinitely.
The list is executed. Stair case wave now consists of 21 steps of .01 sec-
onds except for the 0V step (location 10) which outputs 0V for 0.03 second.
Returns error -221,Settings Error because a LIST command cannot be exe-
cuted while the LIST is still being executed.
Stops execution of the list
Unit outputs the staircase and then the sine wave.
Unit stops generating the waveform.
USING LIST COMMANDS AND QUERIES (SHEET 2 OF 2)
Long Form: LIST:COUNt:SKIP nn
Long Form: LIST:COUNt:SKIP?
LIST:COUN:SKIP
LIST:COUN:SKIP?
BOP-1K 080709