4.1

GENERAL

This section contains the calibration instructions for the Power Supply. It is recommended that
the user be familiar with Local Mode operation (PAR.3.2) before calibrating the unit.
A full calibration consist of a voltage calibration and a current calibration. Both voltage and cur-
rent calibrations consist of zero, max and min, and protection limit calibration. In addition, the
control signals required for series and parallel operation of multiple units are also calibrated. It is
recommended that the unit be recalibrated annually.
In order to enter the calibration mode the correct password may be required (see PAR. 3.2.4.4).
If the password has been lost, call the factory and a special password will be provided. During
the calibration, new calibration data is computed which is then stored in the nonvolatile memory.
Calibration of the BOP is performed in one of three ways: locally using the front panel keys
(PAR. 4.4), remotely by sending commands through either the GPIB or RS 232 bus (PAR. 4.3),
or remotely using the graphical interface supplied with the instrument driver available at
www.kepcopower.com/drivers.htm (PAR. 4.4.3). This VISA compliant driver works with many
GPIB cards from suppliers like National Instruments and Hewlett-Packard. These three ways
cannot be combined. Separate calibration procedures are provided for units connected in series
or parallel.
NOTE: During calibration, various circuits of the BOP are verified. If an error occurs during the
verification process, the unit will beep and an error 24(x), Hardware, (see PAR. B.130)
will be placed in the error queue.
Type
Step
VOLTAGE
1. ZERO
INTERNAL
2. POSITIVE
3. NEGATIVE
VOLTAGE
1. POSITIVE
PROTECTION
2. NEGATIVE
VOLTAGE
1. ZERO
EXTERNAL
2. POSITIVE
VOLTAGE
1. ZERO
GAIN
2. POSITIVE
EXTERNAL
1. ZERO
CONTROLS,
2. POSITIVE
SERIES
BOP-1K 080709

SECTION 4 - CALIBRATION

TABLE 4-1. CALIBRATION SUMMARY
Reference Type
Monitored
and Value
Parameter
Internal
Output Voltage
1. 0.0v
2. +Voltage FS
3. –Voltage FS
Internal
Output Voltage
1. +Voltage FS
2. –Voltage FS
External:
Output Voltage
1. 0.0V
1. 0.0v
2. +10.0V
2. –Voltage FS
(Analog I/O Port,
pin 11 ref to pin 4)
External:
Output Voltage:
1. 0.0V
1. 0.0V
2. +10.0V
2. 0.0V
(Analog I/O Port
pin 11 ref to 4)
External:
Output Voltage:
1. 0.0V
1. 0.0V
2. -10.0V
2. +10.0V
(Par/Ser CTRL-IN
Port, pin 8 ref to 1)
Output Conditions
NO LOAD
1. DVM between OUT S
AUTOMATIC VM
and COM S.
2. Adjust for zero, posi-
tive and negative FS.
NO LOAD
1. DVM between OUT S
AUTOMATIC CM
and COM S.
2. Adjust for positive
and negative FS.
NO LOAD
1. DVM between OUT S
AUTOMATIC VM
and COM S.
2. Adjust for zero, posi-
tive FS.
No Load
1. DVM between OUT S
Automatic VM
and COM S
2. Adjust for zero
(both Zero and Posi-
tive).
No Load
1. DVM between OUT S
Automatic VM
and COM S
2. Adjust for zero,
+10.0V
3. Calibrate unit for
Slave in Series
Notes
4-1