B.71 [SOURce:]LIST:SET:TRIGger? QUERY
Syntax:
Short Form: LIST:SET:TRIG?
Returns <value> = trigger pulse-width between 0.00025 and 0.034 second
Description: Returns trigger pulse-width. If trigger pulse duration has not been set, error 100- "command error"
.
results
B.72 [SOURce:]LIST:SET:WAIT COMMAND
Syntax:
Short Form: LIST:SET:WAIT value
Long Form: LIST:SET:WAIT value
where:
<value> = time-out for a wait command between .00025 and .034 seconds
Description: Establishes the time-out for a wait command. If this command is not sent, the time out period will
be 0 indicating an infinite wait time. (See Figures B-6 and B-7.)
B.73 [SOURce:]LIST:SET:WAIT? QUERY
Syntax:
Short Form: LIST:SET:WAIT?
Returns <value> = time-out for a wait command between 0.00025 and 0.033 second
Description: Returns the time-out for a wait command. Returns 0 if LIST:SET:WAIT not sent, indicating infinite wait
time.
B.74 [SOURce:]LIST:TRIGger COMMAND
Syntax:
Short Form: LIST:TRIG <value>
Long Form: LIST:TRIGger <value>
where:
<value> = programmed current (Amps) for current list, or programmed voltage (Volts) for voltage list.
Description: Similar to LIST:VOLT or LIST:CURR, except also allows Output Off flag of External Protection
Port to be used to trigger an external device. In addition to setting the output to the <value> speci-
fied (for duration established by LIST:SET:TRIG), causes the Output Off flag transistor (EXT_E, pin 5,
and EXT_C, pin 6, of the External Protection port) to be set to the state (on or off) for the duration set
by the LIST:SET:TRIG command.
If the trigger duration is not established by the LIST:SET:TRIG command, a -221 "Settings conflict"
error results. This command must be preceded by LIST:CURR or LIST:VOLT command, otherwise a -
221 "Settings conflict" error results. The entries on the dwell list must match the values on the current
or voltage list, otherwise a -226 "Lists Not Same Length" error results. (See Figure B-7.)
B.75 [SOURce:]LIST:VOLTage COMMAND
Short Form: LIST:VOLT[:LEV] <exp_value>, <exp_value>, . . . (to max of 5900 data points)
Long Form: LIST:VOLTage[:LEVel] <exp_value>, <exp_value>, . . . (to max of 5900 data points)
<exp_value> = digits with decimal point and Exponent, e.g., 2.71E1 for 27.1
Description: Adds the voltage value (in Amps) to list. This command sequentially adds LIST:VOLTage values to
the main channel List Data Table locations illustrated in Table B-2. LIST:CLE sets starting location to 0.
Next location indicated by LIST:VOLT:POIN? The maximum number of entries is 2950, 3933, or 5900.
Since the input buffer of the BOP has a limit of 253 characters, multiple commands are necessary to
complete the full 5900 entries of the list. If LIST:CURR has any entries, an error message: -221,"Set-
tings conflict" is posted in the error queue. Related Commands:
LIST:CLE, *RST. (See example, Figure B-5.)
B.76 [SOURce:]LIST:VOLTage? QUERY
Syntax:
Short Form: LIST:VOLT?
Return Value: <value1>, <value2>, . . . to <value16>
Description: Identifies the parameters (main channel) entered for the list. Starting at location established by
LIST:QUERy, returns comma-separated list of up to 16 values indicating the main channel parameters
entered. i.e., the contents of the main channel locations of Table B-2. Related Commands: LIST:
VOLT, LIST:QUERy. If LIST:CURR has any entries, an error message: -221,"Settings conflict" is
posted in the error queue. (See example, Figure B-5.)
BOP-1K 080709
Long Form: LIST:SET:TRIGger?
Long Form: LIST:SET:WAIT?
Long Form: LIST:VOLTage?
LIST:SET:TRIG?
LIST:SET:WAIT
LIST:SET:WAIT?
LIST:TRIG
LIST:VOLT
LIST:VOLT:POIN?, FUNC:MODE,
LIST:VOLT?
B-19