ERROR MESSAGE
-221,"Settings Conflict"
-222,"Current, Voltage or Data out of
range"
-223,"Too Much Data"
-226,"Lists not same length"
-240,"Hardware error"
-249,"Master Not Found"
-311,"Memory Error"
-314,"Save/recall Memory Lost"
-340,"Calibration Failed"
-350,"Queue Overflow"
-363,"Input buffer overrun"
-400,"QueryError"
-420,"Query Unterminated"
(1)
The Device error bit may be set when the status monitoring functions of the power supply detect an overvoltage/under-
voltage condition.
B-36
TABLE B-5. ERROR MESSAGES (CONTINUED)
ESR ERROR BIT SET
(SEE PAR. A.5)
Execution error bit 4
Execution error bit 4
Execution error bit 4
Execution error bit 4 During a LIST command, number of DWEL list entries was
Execution error bit 4 Power supply did not respond to command.
Execution error bit 4 The slave controller for a multiple unit configuration has not
Device Error bit 3
Device Error bit 3
Device Error bit 3
Device Error bit 3
Device Error bit 3
Query Error bit 2
Query Error bit 2
EXPLANATION
Calibration state not enabled but CALibrate command
received
Value (current or voltage) exceeds power supply rating or
(data) exceeds acceptable command parameters
During a LIST command, the list became full, preventing all
the data from being added to the list
not equal to 1 and did not match number of LIST:VOLT or
LIST:CURR entries.
detected a master unit. - this is an operator error as the
slave should not be operated over the GPIB interface.
(1)
*SAV (save), *RCL (recall), MEMory:SAVE,
MEMory:UPDAte or CALibrate:STORe error
(1)
A memory location that was saved earlier was lost due to a
flash error in Analog processor memory.
(1)
The calibration results were invalid. This normally indicates
an incorrect calibration caused by operator error such as
performing a current calibration without a load,8 or a voltage
calibration with a load.
(1)
Error queue was full, error events have been lost.
(1)
The input from the controlling computer was longer than 253
characters in length.
Data requested was lost due to 253 character limit or BIT
4886 output buffer
Controller sent query but did not accept data from power
supply. See VOLT? (PAR. B.85), CAL:SAVE (PAR. B.85)
commands