Standby Replenishment
Standby replenishment compensates for exhaustion of the developer
due to oxidation and depletion of the fixer due to evaporation.
The replenishment pump releases 100 cc. of chemistry each time it is
activated. You specify the amount of replenishment required for an
entire day. The processor calculates the interval required between each
pulse of fresh chemistry.
Standby replenishment is active continuously, as long as the proces-
sor is plugged in to the power supply. Replenishment continues even
if you shut the processor off with the ON/OFF function key.
Follow this procedure to change the standby replenishment rate:
Step 1:
Press DEV and STDBY REPL. The current standby replen-
ishment rate appears. For example:
Step 2:
Press PRESET.
Step 3:
Press plus (+) or minus (-) to change the setting.
Step 4:
Press Enter to store the new value.
To change the standby replenishment rate for fixer, press FIX instead
of DEV in Step 1.
The fixer replenishment rate needs to be set high enough to maintain
the level in the fix tank.
Range for Fixer STDBY Repl.: 0 - 2000 ml
Range for Developer STDBY Repl.: 0 - 2000 ml
9–16 • Avantra User Guide
DEV STDBY 1500 CC