Lin Triggering - Agilent Technologies InfiniiVision 3000 User Manual

Infiniivision 3000 x-series
Hide thumbs Also See for InfiniiVision 3000:
Table of Contents

Advertisement

12 Press the Standard softkey; then, turn the Entry knob to select the LIN
13 Press the Sync Break softkey and select the minimum number of clocks

LIN Triggering

To set up the oscilloscope to capture a LIN signal, see
Signals"
LIN triggering can trigger on the rising edge at the Sync Break exit of the
LIN single- wire bus signal (that marks the beginning of the message
frame), the Frame ID, or the Frame ID and Data.
A LIN signal message frame is shown below:
1 Press [Trigger].
2 In the Trigger Menu, press the Trigger softkey; then, turn the Entry knob
Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide
One Bit
60%
70%
Sample Point
80%
standard you are measuring (LIN 1.3 or LIN 2.0).
For LIN 1.2 signals, use the LIN 1.3 setting. The LIN 1.3 setting assumes
the signal follows the "Table of Valid ID Values" as shown in section A.2
of the LIN Specification dated December 12, 2002. If your signal does
not comply with the table, use the LIN 2.0 setting.
that define a sync break in your LIN signal.
on page 343.
Sync
Sync
Identifier
Break
Field
Break
Sync Break Exit
to select the serial slot (Serial 1 or Serial 2) on which the CAN signal
is being decoded.
CAN/LIN Triggering and Serial Decode
Data
Checksum
Fields
Field
22
"Setup for LIN
345

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents