To Set Or Clear Persistence - Agilent Technologies InfiniiVision 3000 User Manual

Infiniivision 3000 x-series
Hide thumbs Also See for InfiniiVision 3000:
Table of Contents

Advertisement

To set or clear persistence

With persistence, the oscilloscope updates the display with new
acquisitions, but does not immediately erase the results of previous
acquisitions. All previous acquisitions are displayed with reduced
intensity. New acquisitions are shown in their normal color with normal
intensity.
Waveform persistence is kept only for the current display area; you cannot
pan and zoom the persistence display.
To use persistence:
1 Press the [Display] key.
2 Press Persistence; then, turn the Entry knob to select between:
Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide
Off — turns off persistence.
When persistence is off, you can press the Capture Waveforms softkey
to perform a single- shot infinite persistence. A single acquisition's
data is displayed with reduced intensity, and it remains on the
display until you clear persistence or clear the display.
Persistence — (infinite presistence) Results of previous acquisitions
are never erased.
Use infinite persistence to measure noise and jitter, to see the
worst- case extremes of varying waveforms, to look for timing
violations, or to capture events that occur infrequently.
Variable Persistence — Results of previous acquisitions are erased after
a certain amount of time.
Variable persistence gives you a view of acquired data that is similar
to analog oscilloscopes.
When variable persistence is selected, press the Time softkey and use
the Entry knob to specify the amount of time that previous
acquisitions are to be displayed.
The display will begin accumulating multiple acquisitions.
8
Display Settings
131