Triggers

Edge Trigger

The Edge trigger type identifies a trigger by looking for a specified edge
(slope) and voltage level on a waveform. You can define the trigger source
and slope in this menu. The trigger type, source, and level are displayed in
the upper- right corner of the display.
1 On the front panel, in the Trigger section, press the [Trigger] key.
2 In the Trigger Menu, press the Trigger softkey, and use the Entry knob
3 Select the trigger source:
4 Press the Slope softkey and select rising edge, falling edge, alternating
144
to select Edge.
Analog channel, 1 to the number of channels
Digital channel (on mixed- signal oscilloscopes), D0 to the number of
digital channels minus one.
External — triggers on the rear panel EXT TRIG IN signal.
Line — triggers at the 50% level of the rising or falling edge of the AC
power source signal.
WaveGen — triggers at the 50% level of the rising edge of the
waveform generator output signal. (Not available when the DC, Noise,
or Cardiac waveforms are selected.)
You can choose a channel that is turned off (not displayed) as the
source for the edge trigger.
The selected trigger source is indicated in the upper- right corner of the
display next to the slope symbol:
1 through 4 = analog channels.
D0 through Dn = digital channels.
E = External trigger input.
L = Line trigger.
W = Waveform generator.
edges, or either edge (depending on the selected source). The selected
slope is displayed in the upper- right corner of the display.
Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide