Runt Trigger
The Runt trigger looks for pulses that cross one threshold but not another.
• A positive runt pulse crosses through a lower threshold but not an
• A negative runt pulse crosses through an upper threshold but not a
To trigger on runt pulses:
1 Press the [Trigger] key.
2 In the Trigger Menu, press the Trigger softkey; then, turn the Entry knob
158
High level
Low level
Positive runt pulse
upper threshold.
lower threshold.
to select Runt.
Negative runt pulse