Measurements
X at Max Y
X at Max Y is the X axis value (usually time) at the first displayed
occurrence of the waveform Maximum, starting from the left- side of the
display. For periodic signals, the position of the maximum may vary
throughout the waveform. The X cursor shows where the current X at Max
Y value is being measured.
See Also
•
To measure the peak of an FFT
1 Select FFT as the Operator in the Waveform Math Menu.
2 Choose Math: f(t) as the source in the Measurement Menu.
3 Choose Maximum and X at Max Y measurements.
Maximum units are in dB and X at Max Y units are in Hertz for FFT.
Count Measurements
•
•
•
•
Positive Pulse Count
The Positive Pulse Count measurement is a pulse count for the selected
waveform source.