Arinc 429 Serial Decode - Agilent Technologies InfiniiVision 3000 User Manual

Infiniivision 3000 x-series
Hide thumbs Also See for InfiniiVision 3000:
Table of Contents

Advertisement

You can use the Zoom mode for easier navigation of the decoded data.
To display ARINC 429 serial decode, see
N O T E

ARINC 429 Serial Decode

To set up the oscilloscope to capture ARINC 429 signals, see
ARINC 429 Signals"
For ARINC 429 triggering set up see
N O T E
To set up ARINC 429 serial decode:
1 Press [Serial] to display the Serial Decode Menu.
2 In the submenu accessed by the Settings softkey, you can use the Base
3 Press the Word Format softkey and specify the word decode format:
Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide
Again, label values are always displayed in octal.
on page 398.
softkey to select between hexadecimal and binary display of the
decoded data.
The base setting is used for data display in both the decode line and in
the Lister.
Label values are always displayed in octal, and SSM and SDI values are
always displayed in binary.
Label/SDI/Data/SSM:
Label - 8 bits.
SDI - 2 bits.
Data - 19 bits.
MIL-STD-1553/ARINC 429 Triggering and Serial Decode
"ARINC 429 Serial Decode"
"ARINC 429 Triggering"
on page 401.
"Setup for
on page 399.
26
401