Triggering on FlexRay Frames

1 Press [Trigger].
2 In the Trigger Menu, press the Trigger softkey; then, turn the Entry knob
3 Press the Trigger softkey; then, turn the Entry knob to select Frame.
4 Press the Frames softkey to open the FlexRay Frame Trigger Menu.
5 Press the Frame ID softkey, and use the Entry knob to select the frame
6 Press the Frame Type softkey to select the frame type:
7 Press the Cyc Ct Rep softkey, and use the Entry knob to select the cycle
8 Press the Cyc Ct Bas softkey, and use the Entry knob to select the cycle
Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide
to select the serial bus (Serial 1 or Serial 2) on which the FlexRay
signals are being decoded.
ID value from All or 1 to 2047.
All Frames
Startup Frames
NULL Frames
Sync Frames
Normal Frames
NOT Startup Frames
NOT NULL Frames
NOT Sync Frames
count repetition factor (2, 4, 8, 16, 32, or 64, or All).
count base factor from 0 through the Cyc Ct Rep factor minus 1.
For example, with a base factor of 1 and a repetition factor of 16, the
oscilloscope triggers on cycles 1, 17, 33, 49, and 65.