Transient Withstand Capability; Maximum Input Voltage At Analog Inputs; Maximum Input Voltage At Digital Channels; Environmental Conditions - Agilent Technologies InfiniiVision 3000 User Manual

Infiniivision 3000 x-series
Hide thumbs Also See for InfiniiVision 3000:
Table of Contents

Advertisement

Transient Withstand Capability

C A U T I O N
CAT I 300 Vrms, 400 Vpk; transient overvoltage 1.6 kVpk
50 input: 5 Vrms Input protection is enabled in 50
disconnect if greater than 5 Vrms is detected. However the inputs could still be
damaged, depending on the time constant of the signal. The 50
functions when the oscilloscope is powered on.
With 10073C 10:1 probe: CAT I 500 Vpk, CAT II 400 Vpk
With N2862A or N2863A 10:1 probe: 300 Vrms
C A U T I O N
±40 V peak CAT I; transient overvoltage 800 Vpk

Environmental Conditions

Environment
Ambient
temperature
Humidity
Altitude
Overvoltage
Category
Pollution Degree
Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide

Maximum input voltage at analog inputs

Maximum input voltage at digital channels

Indoor use only.
Operating 0 °C to +55 °C; non-operating –40 °C to +71 °C
Operating: Up to 80% RH at or below +40 °C. Up to 45% RH up to +50 °C.
Non-operating: Up to 95% RH up to +40 °C. Up to 45% RH up to +50 °C.
Operating and non-operating to 4,000 m (13,123 ft)
This product is intended to be powered by MAINS that comply to Overvoltage
Category II, which is typical of cord-and-plug connected equipment.
The InfiniiVision 2000/3000 X-Series oscilloscopes may be operated in
environments of Pollution Degree 2 (or Pollution Degree 1).
Reference
mode and the 50
load will
input protection only
21
317