Replacing The Agilent E8404A Power Supply - Agilent Technologies VXI 75000 Series User And Service Manual

C-size vxi mainframes
Table of Contents

Advertisement

Replacing the
Agilent E8404A
Power Supply
Figure 5-6. Removing the Agilent E8404A Power Supply
Chapter 5
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
1. Turn off the mainframe and remove the power cord.
2. Remove the rear panel. Refer to "Removing the Rear Panel from the
Mainframe" on page 209 for instructions.
3. Using the rings on the power supply, gently pull the supply out from
the plastic connectors on the backplane adapter board. Do not carry
the supply using the rings.
4. Install the exchange power supply. Keep the replacement power
supply to the far right in the mainframe. Make sure the supply is
firmly inserted into the backplane connectors.
5. Replace the rear panel. Make certain that all 14 screws are secured.
Note
You must recalibrate the Enhanced Monitor after replacing the power
supply. Refer to Chapter 4 for calibration details.