STATus:QUEStionable:CURRent:ENABle <mask>
STATus:QUEStionable:CURRent:ENABle?
Parameters
Returned Data
Comments
Error Conditions
Reset Condition
Related
Commands
102 Programming the Enhanced Monitor
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
The
STATus:QUEStionable:CURRent:ENABle
enable register for the CURRent status group.
The
STATus:QUEStionable:CURRent:ENABle?
representing the value of the enable register for the CURRent status group.
The enable register specifies which bits in the event register can generate a summary
bit. The instrument logically ANDs corresponding bits in the event and enable
registers, and ORs all the resulting bits to obtain a summary bit. Summary bits are,
in turn, recorded in another register. The Questionable Current Summary bit is
recorded in the Questionable Condition Register. Enable registers are read-write.
Enable registers are not affected by *CLS (Clear status). Querying enable registers
does not affect them.
Name
Type
Range
<mask>
int16
0 - 32767
Type
Range
int16
0 - 32767
mask value.
Refer to
STATus:QUEStionable:CURRent:CONDition?
each bit in the CURRent status group.
If *PSC was set to 0 prior to power-down, then at power-on, the enable
registers are loaded from non-volatile memory. Therefore, any values that you
have stored (SYSTem:NVSave command) are returned to the registers.
However, if *PSC was set to 1, then at power-on the factory defaults are
returned.
The following table lists the most common error conditions and causes. Error
numbers and corresponding messages can be found using
Number
Message
-222
"Data out of range"
-241
"Hardware missing"
*RST resets register the last value saved by the SYST:NVS command. Use
to clear the enable registers of the OPERation status group and the
STATus:PRESet
QUEStionable status group.
STATus:QUEStionable:CURRent:CONDition?,
STATus:QUEStionable:CURRent:EVENt?, STATus:PRESet, SYSTem:NVSave,
*PSC
command sets the value of the
query returns an integer
Default
487
Bit mask for the enable register of the
(see comments
CURRent status group.
below)
Description
Probable Cause
The <mask> value is invalid.
Power supply doesn't support current measurements
Description
for a description of
query.
SYSTem:ERRor?