STATus:PRESet

Comments
Reset Condition
Related
Commands
94 Programming the Enhanced Monitor
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
The
command sets the enable registers. The Operation and
STATus:PRESet
Questionable Enable Registers are preset to 0, disabling all events. All other enable
registers are preset to 1s (except P5STBY and P5EXT in the
STAT:QUES:VOLT:ENABle
transition filters are preset to 1s.
After executing this command, none of the events in the OPERation event
register or the QUEStionable event register will be reported as a summary bit
in the Status Byte.
This command does not clear or change any of the Events in the OPERation
status group or the QUEStionable status group.
If *PSC was set to 0 prior to power-down, then at power-on, the enable
registers are loaded from non-volatile memory. Therefore, any values that you
have stored (
SYSTem:NVSave
However, if *PSC was set to 1, then at power-on the factory defaults are
returned. The factory defaults are listed in the following table:
STATus:OPERation:ENABle
STATus:QUEStionable:ENABle
STAT:QUES:BLOWer:ENABle
STAT:QUES:CURRent:ENABle
STAT:QUES:TEMPerature:ENABle
STAT:QUES:VOLT:ENABle
STAT:QUES:VOLT:PTR
*RST does not affect the status system.
STATus:OPERation:ENABle, STATus:QUEStionable:ENABle,
STATus:QUEStionable:BLOWer:ENABle,
STATus:QUEStionable:CURRent:ENABle,
STATus:QUEStionable:TEMPerature:ENABle,
STATus:QUEStionable:VOLTage:ENABle
register), enabling all events. The SCPI positive
command) are returned to the registers.
Register
Default
Value
0
0
7
487
32767
487
511