Status:questionable:power:level? <Supply>[,Min|Max] - Agilent Technologies VXI 75000 Series User And Service Manual

C-size vxi mainframes
Table of Contents

Advertisement

STATus:QUEStionable:POWer:LEVel? <supply>[,MIN|MAX]
Parameters
Returned Data
Comments
Error Conditions
Reset Condition
Related
Commands
108 Programming the Enhanced Monitor
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
The
STATus:QUEStionable:POWer:LEVel?
representing the power levels of each power supply in watts.
Name
Type
Range
<supply>
enum.
P24, P12, P5
N2, N5P2, N12
N24, TOTal
optional
optional
MIN, MAX
parameter
enum.
Type
Range
float32
This command is not supported for all power supplies. If the power of the
mainframe is off, this query will return the last measurement made before
power turned off.
The following table lists the most common error conditions and causes. Error
numbers and corresponding messages can be found using
Number
Message
-241
"Hardware missing"
*RST does not affect the status system.
STATus:QUEStionable:CONDition?, STATus:QUEStionable:EVENt?,
STATus:QUEStionable:ENABle?, STATus:QUEStionable:ENABle
query returns a floating number
Default
none
Indicates which power supply wattage is
being queried. P5 is positive 5 Vdc
supply, N12 is negative 12 Vdc supply,
etc.
none
When added after <param>, returns the
maximum and minimum wattage limits.
Default
none
Power level of <supply>, in Watts.
Probable Cause
Power supply doesn't support power measurements.
Description
Description
query.
SYSTem:ERRor?