Status:questionable:voltage:condition - Agilent Technologies VXI 75000 Series User And Service Manual

C-size vxi mainframes
Table of Contents

Advertisement

STATus:QUEStionable:VOLTage:CONDition?

Returned Data
Comments
Reset Condition
Related
Commands
Chapter 3
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
The
STATus:QUEStionable:VOLTage:CONDition?
representing the contents of the condition register associated with the VOLTage
status group.
The condition register continuously monitors the hardware and firmware status of
the instrument. There is no latching or buffering for this register; it is updated in real
time. Condition registers are read-only.
The following table describes each bit in the VOLTage status group:
Bit
0
+24 (P24) volt power supply
1
+12 (P12) volt power supply
2
+5 (P5) volt power supply
3
5Volt Standby
4
External 5 volt supply for
Enhanced Monitor RS-232 port.
5
-2 (N2) volt power supply
6
-5.2 (N5Pt2) volt power supply
7
-12 (N12) volt power supply
8
-24 (N24) volt power supply
9
10
11
12
13
14
15
Type
Range
Default
int16
0 - 32767
none
A bit is set when the supply is out of limits.
*RST does not affect the status system.
STATus:QUEStionable:VOLTage:EVENt?,
STATus:QUEStionable:VOLTage:ENABle
query returns an integer
Description
Unused
Unused
Unused
Unused
Unused
Unused
Unused
Description
Bit mask indicating which conditions are set (1) or cleared (0)
for the VOLTage status group.
Programming the Enhanced Monitor
Decimal Value
1
2
4
8
16
32
64
128
256
512
1024
2048
4096
8192
16384
32768