Test:memory - Agilent Technologies VXI 75000 Series User And Service Manual

C-size vxi mainframes
Table of Contents

Advertisement

TEST:MEMory?

Returned Data
Error Conditions
Reset Condition
Related
Commands
Chapter 3
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
performs a test of the ROM, RAM, VXI communication, and
TEST:MEMory?
non-volatile memory in the enhanced monitor. The command returns an integer;
refer to TEST:ALL? for a complete list of test failure messages.
Type
Range
int16
-32767 to
Result of test. +0 is a successful test. Refer to TEST:ALL? for a
32767
complete list of test failure messages.
The following table lists the most common error conditions and causes. Error
numbers and corresponding messages can be found using
Number
Message
-221
"Settings Conflict"
*RST aborts any TEST being performed.
*TST?, TEST:ALL?, TEST:SENS?, TEST:TEMP?, TEST:DISP?, TEST:BLOW?,
TEST:TIME?
Description
SYSTem:ERRor?
Probable Cause
The mainframe's power is off.
Programming the Enhanced Monitor
query.
161

Advertisement

Table of Contents
loading

This manual is also suitable for:

Vxi e8402aVxi e8404a

Table of Contents