History:time:ltst - Agilent Technologies VXI 75000 Series User And Service Manual

C-size vxi mainframes
Table of Contents

Advertisement

HISTory:TIME:LTST?

Returned Data
Comments
Reset Condition
Related
Commands
Chapter 3
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
returns the amount of operating time that has passed since the
HISTory:TIME:LTST?
last
? or
command was performed.
*TST
TEST
Type
<hours>
uint32
0 to 4294967295
<min>
uint16
0 to 60
<sec>
uint16
0 to 60
This query returns 4294967295,+0,+0 if a
The history queue contains information on which test was done.
*RST has no effect on this query.
*TST?, TEST[:ALL]?, TEST:BLOWer?, TEST:SENSe?, TEST:DISPlay?,
TEST:TEMPerature?, TEST:MEMory?, TEST:TIME?
Range
Default
none
none
none
or
*TST?
Programming the Enhanced Monitor
Description
Hours since *TST
+ minutes since *TST
+ seconds since *TST
has not been done.
TEST