STATus:QUEStionable:CURRent:CONDition?

Returned Data
Comments
Error conditions
Reset Condition
Related
Commands
Chapter 3
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
The
STATus:QUEStionable:CURRent:CONDition?
representing the contents of the condition register associated with the CURRent
status group.
The condition register continuously monitors the hardware and firmware status of
the instrument. There is no latching or buffering for this register; it is updated in real
time. Condition registers are read-only.
The following table describes each bit in the CURRent status group:
Bit
0
+24 (P24) volt power supply
1
+12 (P12) volt power supply
2
+5 (P5) volt power supply
3
Unused
4
Unused
5
-2 (N2) volt power supply
6
-5.2 (N5P2) volt power supply
7
-12 (N12) volt power supply
8
-24 (N24) volt power supply
9
Unused
10
Unused
11
Unused
12
Unused
13
Unused
14
Unused
15
Unused
Type
Range
Default
int16
0 - 32767
none
A bit is set when a current measurement is outside of the user-set limit.
The following table lists the most common error conditions and causes. Error
numbers and corresponding messages can be found using
Number
Message
-241
"Hardware missing"
*RST does not affect the status system.
STATus:QUEStionable:CURRent:EVENt?,
STATus:QUEStionable:CURRent:ENABle
query returns an integer
Description
Description
Bit mask indicating which conditions are set (1) or cleared (0)
for the CURRent status group.
Probable Cause
Power supply doesn't support current measurements
Programming the Enhanced Monitor
Decimal Value
1
2
4
8
16
32
64
128
256
512
1024
2048
4096
8192
16384
32768
query.
SYSTem:ERRor?