TEST:RESults[:CODE]?

Returned Data
Comments
Reset Condition
Related
Commands
162 Programming the Enhanced Monitor
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
returns the results code for the last executed TEST, *TST,
TEST:RESults[:CODE]?
or calibration command. The command returns an integer; refer to TEST:ALL? for
a complete list of test failure messages and CAL:ALL? for a complete list of
calibration failure commands.
Type
Range
int16
-32768 to 32767 Result of test. +0 is a successful test. Refer to TEST:ALL? for a
complete list of test failure messages.
The history queue also contains an entry for each test run and its results and
time stamp. Refer to
Two other codes can be returned:
-- 478 indicating the last calibration or test was aborted.
-- 995 indicating that no calibration or test has been done since mainframe
power-on.
*RST does not change the last test results.
TEST:RES:VERB?
*TST?, TEST:ALL?, TEST:SENS?, TEST:TEMP?, TEST:DISP?, TEST:BLOW?,
TEST:TIME?, HIST:QUEue:FETCh?
Description
HIST:QUE:FETC?