System:communicate:serial[:Receive]:Parity[:Type] <Parity> System:communicate:serial[:Receive]:Parity[:Type] - Agilent Technologies VXI 75000 Series User And Service Manual

C-size vxi mainframes
Table of Contents

Advertisement

SYSTem:COMMunicate:SERial[:RECeive]:PARity[:TYPE] <parity>
SYSTem:COMMunicate:SERial[:RECeive]:PARity[:TYPE]?
Parameters
Returned Data
Comments
Error Conditions
Reset Condition
140 Programming the Enhanced Monitor
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
SYSTem:COMMunicate:SERial[:RECeive]:PARITY
communications. Parameters are EVEN (received parity maintains even parity),
ODD (received parity maintains odd parity), or NONE (no parity bit will be
received). Default is NONE (no parity).
SYSTem:COMMunicate:SERial[:RECeive]:PARITY?
the parity for RS-232 communication. The string is either "EVEN", "ODD", or
"NONE."
Name
Type
Range
<parity> enum EVEN, ODD, NONE NONE
Type
Range
enum EVEN, ODD, NONE RS-232 Parity
While this command operates independently of either the ...BITs and ...SBITs
commands, there are combinations which are not allowed because of their data
frame bit width. The following table shows the possible combinations:.
..BITS
...PARity <type>
7
7
7
ODD/EVEN
7
ODD/EVEN
8
8
8
ODD/EVEN
8
ODD/EVEN
The following table lists the most common error conditions and causes. Error
numbers and corresponding messages can be found using
Number
Message
-222
"Data out of range"
*RST resets parity to the last value saved by the SYST:NVS command.
Default
RS-232 Parity.
Description
...SBITs
NONE
NONE
NONE
NONE
Probable Cause
You tried to set a disallowed combination of Bits,
Parity, and Stop Bits.
sets the parity for the RS-232
returns a string representing
Description
Frame Bits
1
disallowed
2
10
1
10
2
11
1
10
2
11
1
11
2
disallowed
query.
SYSTem:ERRor?