TEST:TIME?

Returned Data
Comments
Error Conditions
Reset Condition
Related
Commands
166 Programming the Enhanced Monitor
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
performs a test of the timing functionality of the enhanced monitor.
TEST:TIME?
This test outputs a waveform on Pin 25 of the Diagnostic Connector with a 16 msec
period for 10 seconds. The waveform will start within 2 seconds of the command.
This waveform can be measured with a counter to determine its accuracy and jitter.
There is no failure return for this test.
This command always returns a +0. Refer to Chapter 4 for instructions on
performing this test.
interrupts the monitoring function of the enhanced monitor.
TEST:TIME?
Measurement levels returned from other commands during a test routine were
all measured before the test routine started.
The following table lists the most common error conditions and causes. Error
numbers and corresponding messages can be found using
Number
Message
-221
"Settings Conflict"
*RST aborts any TEST being performed.
*TST?, TEST:ALL?, TEST:SENS?, TEST:TEMP?, TEST:DISP?, TEST:BLOW?,
TEST:MEM?
SYSTem:ERRor?
Probable Cause
The mainframe's power is off.