HISTory:CURRent:MAXimum? <supply>
Parameters
Returned Data
Comments
Error Conditions
Reset Condition
Related
Commands
Chapter 3
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
HISTory:CURRent:MAXimum?
the maximum amperage that has occurred to the <supply> since manufacture of the
mainframe, or the most recent
Name
Type
<supply>
enum
P5, P12, N12, P24,
N24, N5PT2, N2
Type
Range
float
•
The maximum amperage tracking can be zeroed by the
commands.
The following table lists the most common error conditions and causes. Error
numbers and corresponding messages can be found using
Number
Message
-221
"Settings Conflict"
-224
"Illegal Parameter"
-311
"Memory error"
A *RST has no effect on the HISTory subsystem.
HISTory:CURRent:CMAXimum?, HISTory:CURRent:[HISTogram]?,
HISTory:RESet:CURRent, HISTory:RESet[:ALL]
<supply> returns a single floating point number for
or
HIST:RES:CURR
HIST:RES:ALL
Range
Default
none
Selects the power supply for the
maximum amperage request. P5 is
positive 5 Vdc supply, N12 is
negative 12 Vdc supply, etc.
Description
Maximum current (in amperes) for specified power supply.
Probable Cause
The power supply doesn't support current
measurements.
The <supply> parameter was not correct.
The historical data's memory is not available.
Programming the Enhanced Monitor
.
Description
HISTory:RESet
query.
SYSTem:ERRor?
67