Calibration Subsystem; Calibration[:All] - Agilent Technologies VXI 75000 Series User And Service Manual

C-size vxi mainframes
Table of Contents

Advertisement

Subsystem
Syntax

CALibration[:ALL]?

Returned Data
200 Calibrating and Verifying Performance
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
The calibration subsystem controls the calibration of the mainframe's monitoring
function.
:CALibration
[:ALL]?
:TEMPerature?
:VALue
:TEMPerature <value>
:TEMPerature?
:VOLTage <supply>,<value>
:VOLTage? <supply>
:VOLTage?
performs all calibration procedures. Calibration is performed
CALibration[:ALL]?
assuming the mainframe is in these conditions: no VXI modules installed in
mainframe, mainframe at a steady state temperature in a constant temperature
environment (fan switch set to FULL for at least ten minutes). This means that this
command should always be issued through the Enhanced Monitor's RS-232
interface. Calibrations performed are:
Calculate A/D offset and gain for each VXI power supply. Externally
measured voltages previously reported to the enhanced monitor through
CAL:VAL:VOLT are used for these calculations.
Calculated A/D offset and gain for each temperature sensor. Externally
measured temperatures previously reported to the enhanced monitor through
CAL:VAL:TEMP are used for these calculations. Flag any sensor varying
more than ±2 degree off expected.
Type
Range
int16
0-32767
+0 indicates a satisfactory calibration.
+1057 to 1069: bad front sensor in slot [result-1057]
+1070 to 1082: bad middle sensor in slot [result-1070]
+1083 to 1095: bad rear sensor in slot [result-1083]
+1096: bad voltage gain calculation on +5V Supply
+1097: bad voltage gain calculation on +12V Supply
+1098: bad voltage gain calculation on -12V Supply
+1099: bad voltage gain calculation on +24V Supply
+1100: bad voltage gain calculation on -24V Supply
+1101: bad voltage gain calculation on -5.2V Supply
+1102: bad voltage gain calculation on -2V Supply

CALibration Subsystem

Complete calibration of enhanced monitor.
Calibrates temperature monitoring function
Sets the calibration temperature.
Returns calibration value
Sets the VXI supply calibration voltage
Returns calibration value
Calibrates voltage monitoring function
Description
Chapter 4