Status:questionable:current[:Event] - Agilent Technologies VXI 75000 Series User And Service Manual

C-size vxi mainframes
Table of Contents

Advertisement

STATus:QUEStionable:CURRent[:EVENt]?

Returned Data
Comments
Error Conditions
Reset Condition
Related
Commands
Chapter 3
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
The
STATus:QUEStionable:CURRent:EVENt?
the value of the event register for the CURRent status group.
The event register latches positive transition events from the current condition
register. A positive transition occurs when a condition makes a transition from
unasserted to asserted (a low to a high state). Bits in the event register are latched,
and once set, they remain set until cleared by a
STATus:QUEStionable:CURRent:EVENt?
buffering; so while an event bit is set, subsequent events corresponding to that bit are
ignored. Event registers are read only.
Type
Range
int16
0 - 32767
Refer to
STATus:QUEStionable:CURRent:CONDition?
each bit in the CURRent status group.
The following table lists the most common error conditions and causes. Error
numbers and corresponding messages can be found using
Number
Message
-241
"Hardware missing"
*RST does not affect the status system, use *CLS to clear all event registers.
STATus:QUEStionable:CURRent:CONDition?,
STATus:QUEStionable:CURRent:ENABle, *CLS
query returns an integer representing
query or *CLS (clear status). There is no
Default
none
Bit mask indicating which events are set (1)
or cleared (0) for the CURRent status group.
Probable Cause
power supply doesn't support current
measurements
Programming the Enhanced Monitor
Description
for a description of
query.
SYSTem:ERRor?
103