TEST[:ALL]?

Returned Data
Comments
Error Conditions
Reset Condition
Related
Commands
158 Programming the Enhanced Monitor
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
performs all the tests in the TEST subsystem except TEST:TIME?. They
TEST:ALL?
are performed in the following order: MEMory, SENSe, TEMPerature, DISPlay,
BLOWer. Any failure causes an immediate error return. The command returns an
integer; refer to the table below for valid responses and meaning.
Type
Range
int16
-32767 to
result of test. +0 is a successful test.
32767
+996
+997
+998
+999
+1000 - 1006 Bad internal ground measurements
+1007 - 1019 Bad front slot temperature measurement
+1020-1032 Bad middle slot temperature measurement.
+1033-1045 Bad rear slot temperature measurement.
+1046
+1050-1056 Bad voltage measurement.
+1103-1306
+1307
+1308
+1309
+1310
+1311
+1312-1381 Bad measurement or ground
+1382-1389 Bad fan voltage
+1390-1397 Bad impeller measurement
+1398-1405 Bad power supply fan 1 measurement.
+1406-1413 Bad power supply fan 2 measurement.
This is a complete test. It can take up to 7 minutes before it returns.
TEST:[:ALL]? interrupts the monitoring function of the enhanced monitor.
Measurement levels returned from other commands during a test routine were
all measured before the test routine started.
Executing TEST:RES:VERB? following a test command, returns a string
describing the result.
The following table lists the most common error conditions and causes. Error
numbers and corresponding messages can be found using
Number
Message
-221
"Settings Conflict"
*RST aborts any TEST being performed.
*TST?, TEST:MEM?, TEST:SENS?, TEST:TEMP?, TEST:DISP?, TEST:BLOW?,
TEST:RESults:VERBose?
Test Result Codes and Description
ROM checksum failure
RAM test failure
VXI communication hardware failure
Non-volatile memory corruption
Bad ambient temperature measurement.
Bad A/D or D/A.
Bad multiplexer.
Bad display contrast
Bad front temperature board
Bad middle temperature board
Bad rear temperature board
SYSTem:ERRor?
Probable Cause
The mainframe's power is off.
query.