Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
performs a test of the A/D, D/A, and multiplexers in the enhanced
TEST:SENSe?
monitor. The command returns an integer; refer to TEST:ALL? for a complete list
of test failure messages.
Type
Range
int16
-32767 to 32767 Result of test. +0 is a successful test. Refer to TEST:ALL? for a
complete list of test failure messages.
interrupts the monitoring function of the enhanced monitor.
TEST:SENSe?
Measurement levels returned from other commands during a test routine were
all measured before the test routine started.
The following table lists the most common error conditions and causes. Error
numbers and corresponding messages can be found using
Number
Message
-221
"Settings Conflict"
*RST aborts any TEST being performed.
*TST?, TEST:ALL?, TEST:MEM?, TEST:TEMP?, TEST:DISP?, TEST:BLOW?,
TEST:TIMe?
Description
SYSTem:ERRor?
Probable Cause
The mainframe's power is off.
query.