TEST:TEMPerature?

Returned Data
Comments
Error Conditions
Reset Condition
Related
Commands
Chapter 3
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
performs a test of the 3 temperature boards containing the 39
TEST:TEMPerature?
slot exhaust air temperature sensors. The command returns an integer; refer to
TEST:ALL? for a complete list of test failure messages.
Type
Range
int16
-32767 to 32767 Result of test. +0 is a successful test. Refer to TEST:ALL? for a
complete list of test failure messages.
TEST:TEMPerature?
monitor. Measurement levels returned from other commands during a test
routine were all measured before the test routine started.
The following table lists the most common error conditions and causes. Error
numbers and corresponding messages can be found using
Number
Message
-221
"Settings Conflict"
*RST aborts any TEST being performed.
*TST?, TEST:ALL?, TEST:SENS?, TEST:MEM?, TEST:DISP?, TEST:BLOW?
Description
interrupts the monitoring function of the enhanced
Probable Cause
The mainframe's power is off.
Programming the Enhanced Monitor
query.
SYSTem:ERRor?
165