Module Error Messages; Solvent Zero Counter; Pressure Above Upper Limit - Agilent Technologies Infinity II SFC Binary 1260 User Manual

Table of Contents

Advertisement

Module Error Messages

Solvent Zero Counter

Error ID: 2055
The error message is triggered if the remaining volume in a solvent bottle falls
below the set limit.
Probable cause
1
2

Pressure Above Upper Limit

Error ID: 2014, 2500
The system pressure has exceeded the upper pressure limit.
Probable cause
1
2
3
4
1260 SFC Binary Pump
Volume in bottle below specified volume.
Incorrect setting.
Upper pressure limit set too low.
Blockage in the flowpath (after the
damper).
Defective damper.
Defective main board.
Error Information

Module Error Messages

Suggested actions
Refill bottles and reset solvent counters.
Make sure the set solvent volume matches the
actual bottle filling and set the shutoff limit to a
reasonable value (e.g. 100 mL for 1 L bottles)
Suggested actions
Ensure the upper pressure limit is set to a value
suitable for the analysis.
Check for blockage in the flow path. The
following components are particularly subject
to blockage: inline filter frit, needle
(autosampler), seat capillary (autosampler),
sample loop (autosampler), column frits and
capillaries with small internal diameters (e.g.
50 µm ID).
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.