Replacing boards / Replacing components
•
Click "Erase FLASHes" and wait until the "ready" message appears.
•
Click "Download File" within the next 120 seconds.
•
The download process lasts approx. 25 minutes.
•
"Ready" is displayed after the successful download.
•
Close the download window by clicking the "Back" button.
Loading the parameters
•
Preparations: see Page 6 - 27.
•
Select the "File" "Load" menu.
•
Select Load with technical Data and load the parameter set, stored on the parameter
disk of the MEMOSKOP service program, into the service program.
•
Click Put to MEMOSKOP, the transmission to the MEMOSKOP starts.
Loading the language file
•
Preparations: see Page 6 - 27.
•
Select the "Language" menu.
•
Insert the disk with the language files in drive A.
•
Select the language that is appropriate for your country.
•
Click "Send this Language to MEMOSKOP", the language file is transferred.
Programming the organ menus
NOTE
•
Program the parameters you recorded from the organ menus on the keyboard or by
means of MEMOSKOP service program.
Programming the MEMOSKOP serial number
•
Call up the technical setup on the MEMOSKOP keyboard.
•
Select "More selections" to display the extended menu.
•
Select "Unit serial number".
•
Enter and save the serial number of the MEMOSKOP Fast.
Programming the licenses (Options)
NOTE
•
Program the 5000/9984; Sub; DCM options according to the licenses available.
•
Call up the technical setup on the MEMOSKOP keyboard.
•
Select "More selections" to display the expanded menu.
Siemens AG
Medical Solutions
If the parameters on the disk were successfully saved, the organ
menu will be restored after the parameters are reloaded and do
not need to be reprogrammed.
The licenses available can be retrieved from the system using the
service PC (see Help file / Information Licenses / Version informa-
tion UNIT ID).
SPR2-250.840.01
Rev. 01
Seite 29 von 32
09.05
CS PS 24