System Installation
1
2
Fig. 3
Polydoros SX 65
Move the power cabinet into its location.
Attach the top of the cabinet to the wall with retaining brackets to prevent the cabinet
from tilting.
If required, the cabinet can also be bolted to the floor.
If required, remove the cover for the corresponding cable inlet on the generator cabinet.
Move cabinets A12 and A13 to their location according to the project plan or the above
example (Fig. 1).
The splash guard has to be installed between the control cabinet A 12 and the shock
wave cabinet A13. If necessary, turn it around.
Install the sidewalls of the cabinets.
Within the scope of validity of the UL regulations (e.g. USA) the metal case should not be
removed due to fire protection.
Install the enclosed screw (1/Fig. 3) with nut (2/Fig. 3) to a cabinet to connect the cabinet.
Then move the second cabinet to its position and tighten with nut (4/Fig. 3) and flat
washer (3/Fig. 3).
p
Move the cabinets to their final position and bolt them to the wall and the floor.

Installing and Connecting the High Voltage Charger

Unpack the charging unit, place it on the rails provided in the shock wave cabinet above
the electronics unit N12 and secure it on the front using 4 screws.
Connect the power line cable at N12 (1 on L1, 2 on N and ge/gn on PE).
Plug in the cable from the N12, K108 to the charger controller on the charger.
Siemens AG
Medical Solutions
4
3
RXL2-120.812.01
Rev. 02
10.04
Page 3 of 14