GE H
EALTHCARE
D
FC091194, R
IRECTION
EVISION
4-3-17
ECG Check
4-3-17-1
Introduction
The ECG capability on this unit, is intended as use as a trigger for measurements, but can also be
viewed on the screen.
4-3-17-2
Parts needed
•
ECG Harness, P/N:16L0026 + P/N:16L0028
•
ECG Pads, (3 pc)
or
•
ECG simulator
4-3-17-3
Preparations
None
4-3-17-4
ECG Check
Table 4-32
Step
1.
Connect the ECG harness to the connector on the front
of the system
2.
Connect the three leads to a ECG simulator,
or
Fasten the three ECG Pads to your body and connect
the three leads to respective ECG Pad
4 - 68
11
ECG Check
Task
Section 4-3 - Functional Checks
V
Expected Result(s)
The unit displays a straight curve along the bottom
edge of the image sector on the screen.
- When connecting, the signal on the screen will be
noisy
- When the connection is completed, a typical clean
ECG signal is displayed.
7 S
M
IVID
ERVICE
ANUAL