To Test The Bandwidth (Oscilloscope); Set Up The Equipment - HP 1660E Series Service Manual

Logic analyzers
Table of Contents

Advertisement

To test the bandwidth (oscilloscope)
Testing the bandwidth verifies the performance of the following specification:
• Bandwidth
This test verifies the bandwidth (dc coupled) of the instrument from dc to 500
MHz.
Equipment Required
Equipment
Signal Generator
Power Sensor
Power Splitter
Cable
Adapter
Cr~ical
Specifications
1 - 500 MHz at approximately 170 mV rms
1- 500 MHz±3% accuracy
Outputs differ by <0.15 dB
Type N{mllm} 24-inch
Type N{m)to BNC
IfI
Recommended
ModellPart
HP 8656B
HP 436/8482A
HP 11667B
HP 11500B
HP 1250-0780
Set up the equipment
Turn on the equipment required and the logic analyzer. Let them warm up for 30
minutes if you have not already done so.