Set Up The Logic Analyzer - HP 1660E Series Service Manual

Logic analyzers
Table of Contents

Advertisement

Testing Performance
To test the trigger sensitivity (oscilloscope)
Set up the logic analyzer
1 Set up the Configuration menu.
a Press the Config key. At the pop up menu, select Scope Channel.
b Select the Input field, then select C1.
c Move the cursor to the Probe field, then use the RPG knob to dial in 1:1.
d Move the cursor to the V/Div field, then use the PRG knob to dial in 400 mV.
e Move the cursor to the Offset field. Set the offset to 0 by typing 0, then pressing the
Select key.
f
Move the cursor to the Coupling field, then press Select. Select 500 / DC.
g Move the cursor to the slDiv field, then use the RPG knob to dial in 5.00 ns.
~(
Scope Chonnel
I
(AUIO&COle)
~
I
Input
It
VI!)
I.,.
I~ ~
ICouPlln g ]
~
(1
l
-lOOm'.'
~~
SMIDCJ~
$it"l"'l~
I
['l5plo~
I
[SomPle
I
DHo ocqulreaot'
0 "
5.00
Il~
~
I
Options
Period
I
Ne::! acquisitIon'
1 ns
"
2 Set up the Display menu.
a Press the Display key.
b Select the Mode/Arm field, then select Average.
c Move the cursor to the Average # field. Enter "8" in the front-panel keyboard, then
press Done.
d Select the Grid field and set it to On.
e In the Waveform selection menu, delete channel 2.
If
channell
is
not inserted, insert
channell.
(scolwJt
Scope ['lsploy
( Auto5cale )
~
I
Mode
1
(Avert,,;]
Iconneol
DotS]
(YJ
Meroqe
6
011
On
s/Dty
I
c:o
(DISPlay
I
[~omfll~
I
Doto acquIred Ill'
" ,
5,00 ns
(I
s
Op\100$
,
Pf>rIO(l
rje~t
IIcqul&lIIOn,
,
",
"
3-92